Reference value range of ultrasonic shear wave elasticity and dispersion imaging for detecting liver viscoelasticity in healthy adults |
Received:April 28, 2021 Revised:July 03, 2021 Click here to download the full text |
Citation of this paper:ZHU Yu-li,BAO Jing-wen,WANG Kun,XU Qing-yue,DONG Yi,WANG Wen-ping.Reference value range of ultrasonic shear wave elasticity and dispersion imaging for detecting liver viscoelasticity in healthy adults[J].Chinese Journal of Clinical Medicine,2021,28(5):869-873 |
Hits: 864 |
Download times: 353 |
Author Name | Affiliation | E-mail | ZHU Yu-li | Department of Ultrasound, Zhongshan Hospital, Fudan University, Shanghai 200032, China | | BAO Jing-wen | Department of Ultrasound, Zhongshan Hospital, Fudan University, Shanghai 200032, China | | WANG Kun | Department of Ultrasound, Zhongshan Hospital, Fudan University, Shanghai 200032, China | | XU Qing-yue | Department of Ultrasound, Zhongshan Hospital, Fudan University, Shanghai 200032, China | | DONG Yi | Department of Ultrasound, Zhongshan Hospital, Fudan University, Shanghai 200032, China | | WANG Wen-ping | Department of Ultrasound, Zhongshan Hospital, Fudan University, Shanghai 200032, China Shanghai Institute of Imaging Medicine, Shanghai 200032, China | puguang61@126.com |
|
Abstract:Objective: To explore the optimal method and the medical reference range of liver shear wave elastography (SWE) and shear wave dispersion (SWD) imaging in healthy adults. Methods: From May to October 2020, 50 healthy people in Zhongshan Hospital, Fudan University were selected, the SWE and SWD data of the liver in 50 healthy subjects were collected. The influence of different measuring positions, different sampling areas, different ages or genders on the SWE and SWD values was analyzed, then the normal liver reference range of healthy adults was determined. Results: The success rate of the S5/S6 section of the right lobe of the liver was 100%, which was higher than that of the maximum section of the left lobe of the liver (76%). The internal consistency and dispersion degree of the measured values of the right lobe were also better than those of the left lobe of the liver. So the right anterior lobe of the liver was the best measuring position. There was no significant difference in SWE and SWD values among different sampling area (10 mm and 20 mm in diameter), age, or gender groups. Conclusions: SWE and SWD technology have a high success rate and good stability. The right anterior lobe of the liver is the best site for measuring. Different sampling areas, different gender, and ages have no influence on the measured value. According to our study, the normal range of SWE and SWD values of the liver in healthy adults were 3.93~8.75 kPa and 8.80~14.99 m/(s·MHz), respectively. |
keywords:ultrasonic shear wave elastography ultrasonic shear wave dispersion imaging methodology reference range |